Maintenance and Performance Checks 5-4
Recommended Test
Equipment
Recommended test equipment for troubleshooting, calibration and
performance checking is listed below. Test instruments other than
those listed may be used only if their specifications equal or exceed
the required characteristics.
Test Instrument & Accessories
Minimum Specification
Universal Counter
100 MHz, 10 ppm accuracy
DMM
ACV, 0.2%; DCV, 0.1% accuracy
Distortion Analyzer
10 Hz to 100 KHz
Spectrum Analyzer
1 GHz bandwidth, 1 KHz resolution
Pulse/Function Generator
20 MHz, 10 ns pulsewidth
Oscilloscope
2 channels, 400 MHz analog bandwidth
Feedthrough termination
50
S
, 1%
BNC Cables
1 meter long, 50
S
impedance
20dB feedthrough attenuator
50
S
, 1%
“T” BNC Connector
50
S
, 1%
Performance
Check Procedures
Use the following procedures to check the Model 3152 against the
specifications. A complete set of specifications is listed in Appendix
A. The following paragraphs show how to set up the instrument for
the test, what the specifications for the tested function are, and what
acceptable limits for the test are. If the instrument fails to perform
within the specified limits, the instrument has to be calibrated or
tested to find the source of the problem.
Frequency Accuracy
Accuracy Specifications: 0.01% of reading to 50.00 MHz
Equipment: Counter
1.
Configure the Model 3152 as follows:
Function:
Square
Frequency:
10 Hz
Amplitude:
1 V
Remote Commands: :RES
:OUTP ON
:FUNC:SHAP SQU
:FREQ 10
:VOLT 1
2.
Set the counter to frequency measurement.
Содержание 3152
Страница 16: ...Getting Started 1 7 Figure 1 1 Segment 1 Sin x x Waveform Figure 1 2 Segment 2 Sine Waveform...
Страница 25: ...Configuring The Instrument 2 6 This page was intentionally left blank...
Страница 63: ...Using The Instrument 3 38 This page was intentionally left blank...
Страница 80: ...SCPI Command Reference 4 17 Figure 4 1 SCPI Status Registers...
Страница 121: ...Specifications A 12 This page was intentionally left blank...