Maintenance and Performance Checks 5-6
Equipment: DMM, 50
S
feedthrough termination
1.
Configure the Model 3152 as follows:
Function:
Sine
Frequency:
1.000 MHz
Amplitude:
1.61 V
Remote Commands: :RES
:OUTP ON
:FUNC SIN
:FREQ 1e6
:VOLT 1.61
:VOLT:OFFS 5
2.
Set the DMM to DCV measurements
3.
Connect the Model 3152 output to the DMM input.
Terminate the output with a 50
S
feedthrough termination.
Set the Model 3152 offset and verify the DMM reading as
follows:
OFFS SETTING
DMM READING
±5.000 V
±5.070 V to ±4.930 V
4.
Change the Model 3152 AMPL setting to 161 mV. Set
VOLT:OFFS and verify the DMM reading as follows:
OFFS SETTING
DMM READING
±500.0 mV
±507.0 mV to ±493.0 mV
5.
Change the Model 3152 AMPL setting to 16.1 mV. Set
VOLT:OFFS and verify the DMM reading as follows:
OFFS SETTING
DMM READING
±50.00 mV
±50.70 mV to ±49.30 mV
Squarewave
Characteristics
Specified Transition Time: 5 ns typical (10% to 90% of amplitude)
Specified Aberration: <5% of amplitude
Equipment: Oscilloscope, 20dB attenuator
1.
Configure the Model 3152 as follows:
Function:
Square wave
Frequency:
1.000 MHz
Amplitude:
10 V
Remote Commands: :RES
:OUTP ON
:FUNC:SHAP SQU
:FREQ 10e6
:VOLT 10
Содержание 3152
Страница 16: ...Getting Started 1 7 Figure 1 1 Segment 1 Sin x x Waveform Figure 1 2 Segment 2 Sine Waveform...
Страница 25: ...Configuring The Instrument 2 6 This page was intentionally left blank...
Страница 63: ...Using The Instrument 3 38 This page was intentionally left blank...
Страница 80: ...SCPI Command Reference 4 17 Figure 4 1 SCPI Status Registers...
Страница 121: ...Specifications A 12 This page was intentionally left blank...