Specifications A-2
Triggering Characteristics
Trigger Input
Impedance
1 k
6
, ±5%
Level
±10 V
Resolution
10 mV
Sensitivity
100 mV rms
Accuracy
±(5% of level + 150 mV)
Maximum Input Voltage
30 V rms
Minimum Pulse Width
20 ns
Slope
Positive or negative going edges, programmable
Modes
Normal
Continuous waveform is generated
Triggered
Each input cycle generates a single output cycle
Internal Trigger
An internal timer repeatedly generates a single
output cycle
Gated
External signal enables generator. First output cycle
synchronous with the active slope of the triggering
signal. Last cycle of output waveform is always
completed
External Burst
Preset number of up to 1 million cycles are stimulated
by an internal, external, or manual trigger
Internal Burst
An internal timer repeatedly generates a burst of
up to 1 million counted output cycles
Delayed Trigger
0 to 2 million clock cycles
Trigger Frequency
External
To 5 MHz or VXIbus trigger
Internal
From 15 µs to 1000 s
Accuracy
±4%, 15 µs to 59 µs;
±2%, 60 µs to 0.99 ms;
±0.1%, 1 ms to 99.9 ms;
±0.01%, 100 ms to 1000 s
Trigger Sources
Trigger functionality is optionally to:
(1) cause one complete cycle of the selected waveform to be
output
(2) gate the waveform output while the trigger signal is
asserted
Software
IEEE-STD-488.2 *TRG, VXI Word Serial trigger
Содержание 3152
Страница 16: ...Getting Started 1 7 Figure 1 1 Segment 1 Sin x x Waveform Figure 1 2 Segment 2 Sine Waveform...
Страница 25: ...Configuring The Instrument 2 6 This page was intentionally left blank...
Страница 63: ...Using The Instrument 3 38 This page was intentionally left blank...
Страница 80: ...SCPI Command Reference 4 17 Figure 4 1 SCPI Status Registers...
Страница 121: ...Specifications A 12 This page was intentionally left blank...