Rotor-Gene Q MDx CE User Manual 02/2022
147
Any unauthorized change to the file will invalidate the Run Signature.
6.9.7
Locked templates
It is currently not possible for the user to create read-only template files using the Rotor-Gene Q
software. However, if desired, it can be specified as a requirement that all runs are performed using
a specific template file. In order to ensure read-only access to this template, it should be stored on
a network drive where users cannot modify data. Users can still run and modify their own profiles,
while the template on a network drive such as this is protected. In order to track which template has
been used, the Rotor-Gene Q software stores the name of the template file that was run. This
information can be accessed by clicking the
Settings
button, which then enables the
Run Settings
window to be displayed. The template information is stored in
Other Run Information
.
6.10
Gain menu
Click on the
Gain
menu to view the
Gain Settings
for the current run. This sets the gain of the
specified channel before a run. Gain settings are retained from the last run. These can be modified
if the run has not yet started or in the initial cycles. Use the up/down arrows next to each text field
to modify the fields. Then click
OK
.
The gain can be changed during the initial cycles. A red line will be drawn in the appropriate
channel showing where the gain was changed. Cycles before the gain change will be excluded
from analysis.