Rotor-Gene Q MDx CE User Manual 02/2022
127
6.8.4
Edit Samples
Click on the
Samples
button to bring up the
Edit Samples
window. The
Edit Samples
window can
also be accessed by right-clicking over the sample list on the right of the screen. This window has
identical functionality to the
Edit Samples
window in the wizards, except that the toolbar functions
are also available in the File and Edit menus.
Four menus appear at the top of the window,
File
,
Edit
,
Format
, and
Security
. The File menu is used
to create a new (blank)
Edit Samples
window, to open an existing sample template, or to save
sample names as a template for future use. The extension of these template files is *
.smp
. The
Edit
menu allows rows to be copied and pasted. The Security menu enables the sample definitions to
be locked.
Note
: If the sample names are entered very fast during the run (e.g., using a bar code scanner) it
can lead to transposed letters within the sample names. Therefore, it is recommended to avoid using
a bar code scanner and, if applicable, enter the sample names after the run has finished.
This drop-down menu is used to choose a suitable format for the concentration
display. Concentrations are automatically formatted according to the currently
selected location.
This drop-down menu sets the units of measurement for the assay.