Liquidew I.S. User Manual
Michell Instruments
47
APPENDIX A
Appendix A
Technical Specifications
Sensors
Sensor Technology
Michell Ceramic Moisture Sensor
Sensor Version
Easidew PRO I.S.
Measurement Range
0.001...1000 ppm
W
Higher range on request
Actual range dependent on solubility of sample fluid
Calibration Range
-100...+20 °Cdp (-148...+68 °Fdp)
Accuracy
Dew point:
Moisture
content:
±1 °C between –59.9 & +20 °Cdp
(±1.8 °F between -75.9...+68 °Fdp)
±2 °C between –60 & -100 °Cdp
(±3.6 °F between -76...-148 °Fdp)
±10% of reading
±20% of reading
Resolution
0.1 °C b20 and -100 °Cdp (+68 and-148 °Fdp)
Temperature Measurement
Pt100
Temperature Measurement
Range
-20...+70 °C (-4...+158 °F)
Temperature Measurement
Accuracy
Accuracy ±0.2 °C
Analysis Pressure
Up to 5 MPa (50 barg / 725 psig)
Analysis Temperature
0...+50 °C (32...+122 °F)
Sample Flow Rate
Min 0.01 l/min (0.003 GPM), Max 10 l/min (2.64 GPM))
0.1...0.3 l/min (0.026...0.079 GPM) recommended
Calibration
Traceable to British (NPL) and American (NIST) National Humidity
Standards
Control Unit
Display
Two line 6-digits LED, displaying moisture content / dew point (user
toggle) and temperature
Analog Output
Two 4...20 mA (max load 500 Ω) - user configured for parameter, unit
and range.
Output 1 = 21 mA when a high or low error occurs.
Output 2 = 3 mA for a low error and 21 mA for a high error.
Digital Output
RS485 Modbus RTU
Display Mode
Moisture content (ppm
W
)
Dew point (°C or °F)
Temperature (°C or °F)
Display Resolution
0.1 °Cdp, 0.1 °Fdp, 0.01 ppm
W
, 0.1 °C temp
Alarms
Four alarm relays. Control action and set-point are user programmable
Two Form C contacts rated 10A / 240 V AC or 8A / 24 V DC Non-
inductive load
Two Form A contacts rated 5A / 240 V AC or 4A / 24 V DC Non-inductive
load
I.S. Barriers
Galvanic isolation type, integrated to Control Unit
Power Supply
85...265 V AC 47/63Hz or 10...72 V DC
10 W max power consumption
Содержание Michell Instruments Liquidew I.S.
Страница 54: ...Liquidew I S User Manual 46 97092 Issue 5 3 September 2021 APPENDIX A Appendix A Technical Specifications ...
Страница 58: ...Liquidew I S User Manual 50 97092 Issue 5 3 September 2021 APPENDIX B Appendix B Serial Communications ...
Страница 63: ...Liquidew I S User Manual Michell Instruments 55 APPENDIX C Appendix C Modbus RTU Comms ...
Страница 65: ...Liquidew I S User Manual Michell Instruments 57 APPENDIX C ...
Страница 66: ...Liquidew I S User Manual 58 97092 Issue 5 3 September 2021 APPENDIX D Appendix D Hazardous Area Certification ...
Страница 69: ...Liquidew I S User Manual Michell Instruments 61 APPENDIX E Appendix E System Drawings ...
Страница 76: ...http www michell com www ProcessSensing com ...