INTRODUCTION
1.5
Sampling System
The Liquidew I.S. requires a clean sample of the process liquid that meets the
temperature, pressure and flow requirements of the transmitter. The design of the
sampling system will depend on the specific application.
The requirements for the sample liquid going into the sensor block are as follows:
•
Temperature: 0...+40 °C (+32...+104 °F) (m60 °C (+140 °F))
NOTE: Sample temperature must be above the moisture saturation
temperature of the sample fluid
•
Maximum pressure: 45 MPa (450 barg / 6526 psig)
•
Flow rate: 0.1...0.3 l/min (0.026...0.079 GPM)
NOTE: Contact Michell Instruments if you wish to order a specific sampling
system.
Please refer to the Liquidew I.S. Premium Sampling System Instructions if a Michell
sampling system has been ordered with the Liquidew I.S.