Liquidew I.S. User Manual
44
97092 Issue 5.3, September 2021
OPERATION
3.9.3 Alarm Source Selection
The alarm source can be also selected in the Alarm Control Setting page.
From the Main Reading Page, press and hold the
Select
key then press the
key to
enter the ‘ALARM SETTING’ page. Press the
key until SPC_1 is shown.
The user can select
RESULT
(moisture),
DP_C
(dew point °C),
TEMP_C
(temperature
°C),
DP_F
(dew point °F) or
TEMP_F
(temperature °F) as the alarm source.
Other
options like DISP and TOT_1 in the list are for factory settings only and should
not be changed
.
a. Use the
and
keys to adjust the control code to
010
0 1 0
S P C _ 1
b. Press the
Select
key to confirm this control code. The display will then
change to the Alarm Source Selection sub menu.
D P _ C
S O U R C E
c. Change the selection by scrolling through these options using the
and
keys.
d. When the required source is shown on the display, press the
Select
key
to confirm. The alarm will then switch to the selected source. The display
will then return to the Alarm Control Setting page.
0 1 0
S P C _ 1
e. Use the
and
keys to change the control code back to
000
or
100
depending on the alarm type (high/low).
f.
Press the
Select
key to confirm. The display will then change to the Alarm
Control Setting page for Alarm Relay 2.
g. Repeat the same operation to adjust the set-point for Alarm Relay 2,
Alarm Relay 3 and Alarm Relay 4 accordingly.
h. After Alarm Relay 4 is configured, press the
Select
key to return to the
Main Reading Page.
CAUTION: The Alarm Set-Point value will NOT be converted automatically
based on the new alarm source selected. In some cases, the operation to
change the alarm source may trigger the alarm.
For example: If the SP_1 is set
at 10ºC dew point - when the alarm source is changed from
DP_C
(ºC dew point) to
DP_F
(ºF dew point) under SPC_1, the value of SP_1 will be 10ºF.
Содержание Michell Instruments Liquidew I.S.
Страница 54: ...Liquidew I S User Manual 46 97092 Issue 5 3 September 2021 APPENDIX A Appendix A Technical Specifications ...
Страница 58: ...Liquidew I S User Manual 50 97092 Issue 5 3 September 2021 APPENDIX B Appendix B Serial Communications ...
Страница 63: ...Liquidew I S User Manual Michell Instruments 55 APPENDIX C Appendix C Modbus RTU Comms ...
Страница 65: ...Liquidew I S User Manual Michell Instruments 57 APPENDIX C ...
Страница 66: ...Liquidew I S User Manual 58 97092 Issue 5 3 September 2021 APPENDIX D Appendix D Hazardous Area Certification ...
Страница 69: ...Liquidew I S User Manual Michell Instruments 61 APPENDIX E Appendix E System Drawings ...
Страница 76: ...http www michell com www ProcessSensing com ...