Liquidew I.S. User Manual
Michell Instruments
35
OPERATION
3.5.2
User-Defined Liquid
If the process liquid is not listed the user can define their own liquid solubility profile
by entering the
C
S
value (10 pre-set temperature points, over the range 0...+60 °C
(32...+140 °F)) for that particular liquid. There are two user-defined liquids available
and these are named USER 1 and USER 2.
a. In the liquid list in the ‘LIQUID SETUP’ sub menu, use the
and
keys
to scroll to:
U S E R
1
U S E R
1
User-defined liquids are found near to the end of the list.
b. Press the
Select
key to access the
C
S
value table and define the solubility
profile of this liquid. The first temperature point for the
C
S
value is at 0 °C.
? ? ?
0 C
C
S
value
Temp point °C
c. Use the
and
keys to adjust the
C
S
value.
d. Press the
Select
key to confirm. The display will then display the next
temperature point, -10 °C.
? ? ?
1 0 C
C
S
value
Temp point °C
e. Follow steps c. and d. to enter the
C
S
value for the 10 °C temperature
point, and then repeat for 20 °C, 30 °C, 40 °C, 50 °C, 60 °C. If
C
S
values are not available for temperature points outside of expected
measurement conditions then enter an estimated value (extrapolated by
drawing a curve in such cases).
f.
After entering the
C
S
value for the 60 °C temperature point, press the
Select
key to confirm. The display will then return to the Main Reading
Page, and calculations will then based on the USER 1 solubility profile.
For the second user-defined liquid USER 2, the solubility profile can be
entered in USER 2, following the above procedure:
U S E R
2
U S E R
2
Содержание Michell Instruments Liquidew I.S.
Страница 54: ...Liquidew I S User Manual 46 97092 Issue 5 3 September 2021 APPENDIX A Appendix A Technical Specifications ...
Страница 58: ...Liquidew I S User Manual 50 97092 Issue 5 3 September 2021 APPENDIX B Appendix B Serial Communications ...
Страница 63: ...Liquidew I S User Manual Michell Instruments 55 APPENDIX C Appendix C Modbus RTU Comms ...
Страница 65: ...Liquidew I S User Manual Michell Instruments 57 APPENDIX C ...
Страница 66: ...Liquidew I S User Manual 58 97092 Issue 5 3 September 2021 APPENDIX D Appendix D Hazardous Area Certification ...
Страница 69: ...Liquidew I S User Manual Michell Instruments 61 APPENDIX E Appendix E System Drawings ...
Страница 76: ...http www michell com www ProcessSensing com ...