Liquidew I.S. User Manual
30
97092 Issue 5.3, September 2021
OPERATION
3.2.2 Powering-up the Analyzer
For the AC power supply version of the Liquidew I.S., turning on the power switch on the
back panel of the control unit will power-up the control unit and the two transmitters.
There is no power switch for the DC power supply version. It is turned on automatically
once 24 V DC power source is applied.
After turning on the analyzer the display will be illuminated. The default main reading
page displays sample temperature (°C) and sample dew point (°C).
It is possible to change the display unit of the Main Reading Page using the
Display
key (see Section 1.4.1). At the default screen, pressing the
Display
key will change the
units of measurement readings in sequence as shown below:
Temperature
o
C
Dew point
o
C
Temperature
o
C
Moisture ppm
W
Temperature
o
F
Temperature
o
F
Moisture ppm
W
Dew point
o
F
Default
D E W P O I N T
3 9. 3 C
-
2. 2 C
3 9. 3 C
2
2.
1 5
3 9. 3 C
2
8. 2
F
3 9. 3 C
M O
I
S T
U
R E
M O
I
S T
U
R E
2
2. 1 5
1 0 2. 7 F
1 0 2. 7 F
1 0 2. 7 F
1 0 2. 7 F
D E W P O I N T
Figure 21
Units of Measurement Readings in Sequence
Содержание Michell Instruments Liquidew I.S.
Страница 54: ...Liquidew I S User Manual 46 97092 Issue 5 3 September 2021 APPENDIX A Appendix A Technical Specifications ...
Страница 58: ...Liquidew I S User Manual 50 97092 Issue 5 3 September 2021 APPENDIX B Appendix B Serial Communications ...
Страница 63: ...Liquidew I S User Manual Michell Instruments 55 APPENDIX C Appendix C Modbus RTU Comms ...
Страница 65: ...Liquidew I S User Manual Michell Instruments 57 APPENDIX C ...
Страница 66: ...Liquidew I S User Manual 58 97092 Issue 5 3 September 2021 APPENDIX D Appendix D Hazardous Area Certification ...
Страница 69: ...Liquidew I S User Manual Michell Instruments 61 APPENDIX E Appendix E System Drawings ...
Страница 76: ...http www michell com www ProcessSensing com ...