Liquidew I.S. User Manual
Michell Instruments
51
APPENDIX B
Appendix B
Serial Communications
To communicate with the monitor:
•
Connect to the serial port using the wiring defined below.
•
Set the address of the monitor using the front panel.
•
Set the desired communication protocol (ASCII or Modbus RTU) via the
front panel or over the serial interface.
•
Determine the register number of the parameter to be read.
•
Send the correct command to the monitor and decode the response.
RS485 Wiring
The Liquidew I.S. monitor uses a 2-wire RS485 connection.
Pin Name
Liquidew DB9 Pin
A
3
B
2
GND
5
Pin numbers in the manual refer to standard pins on the DB9 D-Sub connector on the rear panel of
the MCU:
Male DB9 Pinout (Liquidew I.S. Monitor on MCU Panel)
1
B A
GND
2
3
4
5
6
7
8
9
It will be necessary to match up the A/B (Differential data pair) and GND (0 V) pins with the wiring
of your own third party adaptor.
E.g. For the K3-ADE RS232->RS485 adaptor use the following wiring:
Pin Name
Liquidew DB9 Pin
K3-ADE DB9 Pin
A
3
8
B
2
3
GND
5
5
This wiring is the generally accepted standard for most 2-wire RS485 equipment.
Содержание Michell Instruments Liquidew I.S.
Страница 54: ...Liquidew I S User Manual 46 97092 Issue 5 3 September 2021 APPENDIX A Appendix A Technical Specifications ...
Страница 58: ...Liquidew I S User Manual 50 97092 Issue 5 3 September 2021 APPENDIX B Appendix B Serial Communications ...
Страница 63: ...Liquidew I S User Manual Michell Instruments 55 APPENDIX C Appendix C Modbus RTU Comms ...
Страница 65: ...Liquidew I S User Manual Michell Instruments 57 APPENDIX C ...
Страница 66: ...Liquidew I S User Manual 58 97092 Issue 5 3 September 2021 APPENDIX D Appendix D Hazardous Area Certification ...
Страница 69: ...Liquidew I S User Manual Michell Instruments 61 APPENDIX E Appendix E System Drawings ...
Страница 76: ...http www michell com www ProcessSensing com ...