INSTALLATION
2.4.2.4 Alarm Output Connection
Four alarm output ports are provided and are connected to the instrument via a single
8-way push fit connector block labelled
ALARMS
as shown below.
ALARMS
socket
Terminal bock
NC
NO
CO
M
CO
M
NC
NO
NO
NO
SP2
SP4
SP1
SP3
Figure 14
ALARM Connector Block
Alarm 1 (connection labelled as SP1) and Alarm 2 (connection labelled as SP2) are Form
C (single pole, double throw) relays.
The method of connection is as follows:
a. Remove the terminal block from the
ALARMS
socket.
b. Strip back the wires of the Alarm 1 cable, exposing approximately
6mm (0.2”) - the use of crimps/wire ferrules is recommended.
c. Insert the
N/O
connection lead into the
SP1
NO
terminal
way on the terminal block and tighten the screw.
d. Insert the
N/C
connection lead into the
SP1
NC
terminal way on the
terminal block and tighten the screw.
e. Insert the common lead into the
SP1
COM
terminal way on the terminal
block and tighten the screw.
f.
Repeat operations b. to e. for connecting the Alarm 2 cable to the SP2
terminals.
Содержание Michell Instruments Liquidew I.S.
Страница 54: ...Liquidew I S User Manual 46 97092 Issue 5 3 September 2021 APPENDIX A Appendix A Technical Specifications ...
Страница 58: ...Liquidew I S User Manual 50 97092 Issue 5 3 September 2021 APPENDIX B Appendix B Serial Communications ...
Страница 63: ...Liquidew I S User Manual Michell Instruments 55 APPENDIX C Appendix C Modbus RTU Comms ...
Страница 65: ...Liquidew I S User Manual Michell Instruments 57 APPENDIX C ...
Страница 66: ...Liquidew I S User Manual 58 97092 Issue 5 3 September 2021 APPENDIX D Appendix D Hazardous Area Certification ...
Страница 69: ...Liquidew I S User Manual Michell Instruments 61 APPENDIX E Appendix E System Drawings ...
Страница 76: ...http www michell com www ProcessSensing com ...