3350F Series Operation Manual 29
When the test is started the right display will show the current value
Being taken by the load. If the Device Under Test is able to supply the
Load according to the values set then the middle display will show
PASS and the right display will show the maximum current taken
During the OCP test. If, during the test, OTP is displayed the over
Temperature protection has been engaged. Similarly if OPP is shown
On the display the over power protection has been activated.
3.2.8. and CC, CR, CV, CP Indicator
There are four operating modes. These can be selected in turn by pressing the
"MODE" key on the 3350F series Electronic Load module. The sequence is:
(CC) Constant Current
(CR) Constant Resistance
(CV) Constant Voltage
(CP) Constant Power
The appropriate LCD will illuminate according to the operating mode is selected.
3.2.9. Key and LED
The input to the 3350F series Electronic Load can be switched ON/OFF by using
The “LOAD” button. Indication of the ON/OFF state is provided by illumination of
The button.
LOAD button lit
= LOAD ON
(load sinks according to the preset values)
LOAD button unlit = LOAD OFF
(the load does not sink current)
Turning the LOAD OFF does not affect the preset values. When the LOAD ON state
Is enabled the unit will revert to sinking according to the preset values.
3.2.9.1. When the Load ON/OFF key is operated the current taken by load will follow
The RISE or FALL with time according to the preset rate. The current RISE
And FALL times can be adjusted in the DYN Setting button of the front panel.
3.2.9.2. In addition to the LOAD ON/OFF function the user can also adjust the
Voltage level at which the unit will automatically start or stop sinking energy.
The adjustable LDon and LDoff voltage levels are found within the CONFIG
Menu. Please note that the LDoff level cannot be set higher than the LDon
Level.
3.2.10. / STA key and LED
The DYN button allows the user to switch between DYNAMIC operation and
STATIC operation. Dynamic operation is only possible in constant current (CC) or
Constant power (CP) mode only. The LED next to the DYN button will become lit
When DYNAMIC operation is selected. If you are in constant resistance (CR) or
Constant voltage (CV) mode pressing the DYN button will have no effect.
MODE
LOAD
DYN