110 PRODIGIT
5.10.1.8. Press START/STOP test key.
5.10.1.9. The UUT’s output voltage drop-out lower than the threshold voltage (V-th
Setting), and the OCP trip point is between I_Hi and I_Lo limitation, then
Middle 5 digits LCD display will shows "PASS", otherwise shows "FAIL".
5.10.2 Remote control OCP
EX
:
REMOTE
(
Set Remote
)
TCONFIG OCP
(
Set OCP test
)
OCP:START 3
(
Set start load current 3A
)
OCP:STEP 1
(
Set step load current 1A
)
OCP:STOP 5
(
Set stop load current 5A
)
VTH 0.6
(
Set OCP VTH 0.6V
)
IL 0
(
Set current low limit 0A
)
IH 5
(
Set current high limit 5A
)
NGENABLE ON
(
Set NG Enable ON
)
START
(
Start OCP testing
)
TESTING?
(
Ask Testing? 1
:
Testing
,
0
:
Testing End
)
NG?
(
Ask PASS/FAIL?
,
0
:
PASS
,
1
:
FAIL
)
OCP?
(
Ask OCP current value
)
STOP
(
Stop OCP testing
)