3350F Series Operation Manual 41
3.2.17. Key and LED
The SHORT key allows the parameters of a SHORT circuit test to be entered. The
SHORT test will attempt to sink high current up to the 3350F load maximum
Current in order to check the power source’s protection and behavior. The test time
can be adjusted and threshold values for the high and low voltage limits set.
Pressing the SHORT key once will cause the button to illuminate. The message
“SHORT PRESS START” will be shown across the 3 displays.
Each press of the SHORT key moves the menu on one step. The left and Middle
LCDs show the currently selected test parameter as text. The value is adjusted by
The rotary knob and can be read from the right display during Setting.
The setting sequence is shown below:
SHORT PRESS START (pressing the red start/stop key starts test)
SHORT TIME (CONTI = Continuous or 100ms to 10,000ms possible)
SHORT V_Hi (High voltage threshold setting)
SHORT V_Lo (Low voltage threshold setting
Exit SHORT test set-up
Press any key to goes to normal mode of LCD display.
Short