3350F Series Operation Manual 95
Chapter 5 Applications
This chapter describes the application information of 3350F Series Electronic Load.
5-1. Local sense connections
Fig 5-1 illustrates a typical set up with the electronic load connected to the DC power supply.
Local sensing is used in application where lead lengths are relatively short, or where load
regulation is not critical.
The 5 digit voltage Meter of 3350F Electronic load measures the voltage of DC INPUT
Terminal automatically; load leads should be bundled or tie-wrapped together to minimize
inductance.
Fig 5-1 Local voltage sense connections
DC Power Supply
- +