3350F Series Operation Manual 109
5-10. Power Supply OCP testing
5.10.1 OCP Manual control
Example:
5.10.1.1. First, press Limit Key function to setting I_Hi 6A.
5.10.1.2. Press Limit Key function to setting I_Lo 0A.
5.10.1.3. Setting OCP test, press OCP key to the next step.
5.10.1.4. Setting start load current 0A, press OCP key to the next step.
5.10.1.5. Setting step load current 0.01A, press OCP key to the next step.
5.10.1.6. Setting stop load current 0.65A, press OCP key to the next step.
5.10.1.7. Setting OCP VTH 0.600V, press OCP key to the next step.