112 PRODIGIT
5.11.1.9. The UUT’s output voltage drop-out lower than the threshold voltage (V-th
Setting), and the OPP trip point is between W_Hi and W_Lo limitation, then
Right 5 digits LCD display will shows "PASS", otherwise shows "FAIL".
5.11.2. Remote control OPP
EX
:
REMOTE
(
Set Remote
)
TCONFIG OPP
(
Set OCP test
)
OPP:START 3
(
Set start load watt 3W
)
OPP:STEP 1
(
Set step load watt 1W
)
OPP:STOP 5
(
Set stop load watt 5W
)
VTH 0.6
(
Set OPP VTH 0.6V
)
WL 0
(
Set watt low limit 0W
)
WH 5
(
Set watt high limit 5W
)
NGENABLE ON
(
Set NG Enable ON
)
START
(
Start OPP testing
)
TESTING?
(
Ask Testing? 1
:
Testing
,
0
:
Testing End
)
NG?
(
Ask PASS/FAIL?
,
0
:
PASS
,
1
:
FAIL
)
OPP?
(
Ask OPP watt value
)
STOP
(
Stop OPP testing
)