14
Cal2Standard
1 character
Indicates whether the ZGM 1120 shall be calibrated on the
provided working standard or on an alternative 2nd
calibration standard.
0 = working standard
1 = 2nd standard
Cal2StandValue
Calibration value for the 2nd standard, in case the Flag
„Cal2Standard“ has been set on 1.
This value is indicated in decimal gloss units (dGU).
Delimiter „|“:
1 character
Separator between single parameters in a command,
respectively, return string.
6.2.5 Serial number (serial No.)
Every ZGM 1120 has a unique 9-digit serial number for identification. The serial number has
to be indicated for the addressing of the requested device in every command-string.
In the additional answer or possible error string this serial number is always mirrored.
NOTE:
The serial number has always to be transmitted with a leading space character.
6.2.6 Transactional number (TID)
The TID is a two digit number, which serves the allocation of the command and answer-string.
It can be generated of two arbitrary ASCII-character (for example xy).
Excluded are only the following 4 ASCII-characters:
„NULL“ (ASCII 0x00)
„:“ (ASCII 0x3A)
-> End of command characters
„A“ (ASCII 0x41)
„|“ (ASCII 0x7C)
-> Parameter delimiter characters
At every command strings, such a TID has to be added. In the additional answer or possible
error string, this TID is always mirrored.
In general, it is not necessary to generate a new TID for every command string. The same TID
can be used all the time.