18. May 2021
© Copyright 2020, PROCEQ SA
21
Parameter:
SerialNo
9-digit serial number with preceding blank
TID
2 characters transaction number of the host
Angle
Indication of the angle to be calibrated
1 = Angle 1 (A1),
2 = Angle 2 (A2),
4 = Angle 3 (A3)
Cal2Standard
Flag, indicating on which calibration standard
the calibration will be carried out:
0 = working standard
1 = 2nd calibration standard
Cal2StandValue
Calibration value of the 2nd calibration
expressed in deci gloss units
(e.g. 95.8 GU => 958).
Return string:
SerialNo
9-digit serial number with preceding blank
TID
2 characters transaction number
deviation
Deviation to the original Zehntner factory
calibration in ppm (points per million),
e.g. 6235 = 0.6235%.
Example:
If you want to calibrate the 85° angle of a three-angle glossmeter (20°/60°/85°) on the
working standard, you need to use the following command string:
72| 401120999|xy|4|0:
Operation code:
72
(Calibrate)
Serial-No:
401120999
(with preceding blank)
TID:
xy
(can be chosen freely)
Angle:
4
(angle A3)
Cal2Standard:
0
(calibration on the calibration standard)