12
a transaction ID (TID) of the performed command
optionally one or more return values
The components of the return string are separated from each other by a
“|“
character (ASCII 0x7C).
In contrast to the command string, the return string is not finished with a
“:“
character (ASCII 0x3A)!
In case there is an error at carrying out the command, an error string will be sent back instead
of a return string (see section 6.5).
6.2.3 Operation code (OP-Code)
The first transmitted component, the so called operation code or OP-code, describes the
command which has to be performed.
The following table shows the assignment of the OP-codes to the corresponding commands:
Command
OP-Code
MeasureValue(SerialNo, TID, AngleBinary, Count, isTemp)
01
GetIsOnStandard(SerialNo, TID)
28
MeasureTemp(SerialNo, TID)
36
ControlLEDOn(SerialNo, TID, LED)
48
ControlLEDOff(SerialNo, TID, LED )
52
Error(SerialNo, TID, ErrorCode) -> Error-String from client to host
56
Reset_Device(SerialNo, TID)
64
Calibrate(SerialNo, TID, Angle, Cal2Standard, Cal2StandValue)
72
AutoSend(SerialNo, TID, AutoSendCluster)
6
Attention:
The listed commands are not the entire set of OP-Codes which the ZGM 1120 Glossmeter can
understand and process. Certain OP-Codes are only used for using the ZGM 1120 with the
Zehntner „GlossTools” software or for factory configuration / calibration.
Only use the operation codes which are listed here. Using unlisted operation codes
could lead to serious manipulations in the device memory, which damages its function
and consequently it has to be sent in for maintenance and calibration to the company
Zehntner.