The corresponding return string could be as follows:
72| 401120999|xy|5361
Operation code:
1
(Calibrate)
Serial-No:
401120999
(with preceding blank)
TID:
xy
(mirrored TID of the command string)
deviation:
5361
ppm (corresponds to 0.5361 % of the original factory calibration)
The value of the calculated deviation to the factory calibration should be smaller than
10 % at a calibration on the calibration standard. If the deviation is higher than 10 %,
please check if the working standard is clean and repeat the calibration. If the deviation
is still higher than 10 %, the ZGM 1120 Zehntner-Glossmeter should be sent to
Zehntner for maintenance and factory calibration.
6.4 Command AutoSend()
If the AutoSend function is enabled the gloss meter will send a measurement return string
when the measure button on the device is pressed.
The result string is sent automatically, without having sent a measure command in advance.
AutoSend(SerialNo, TID, AutoSendCluster)
OP-Code: 6
Parameter:
SerialNo
9-digit serial number with preceding blank
TID
2 characters transaction number of the host
AutoSendCluster 3 characters
AutoSendCluster:
1
st
character
Enables the AutoSend function
1
= AutoSend enabled
0
= AutoSend disabled
2
nd
character
Indication of the angle(s) that has to be measured
Binary coded integer number (see table below)
Integer
A3
Binary Digit 2
2
A2
Binary Digit 2
1
A1
Binary Digit 2
0
1
0
0
1