18. May 2021
© Copyright 2020, PROCEQ SA
16
isTemp
Flag indicates if the temperature is measured while measuring
0 = without temperature
1 = with temperature
Return string:
SerialNo
9-digit serial number with preceding blank
TID
Transactional number [2 characters]
Measuring value
Measurements in „deci gloss units” (dGU).
(900 dGU would mean 90 gloss units)
If an angle is not measured, a „
-
1“ is returned
[max. 3 x 5 characters]
Offset
Offset respectively value of the zero current measurement (raw
value of the A/D-modifier).
In case an angle is not being measured, „
-
1“ is returned for the
corresponding offset value.
If an „Overflow“occurs during a measurement, a „
-
2“ is returned.
[max. 3 x 5 characters]
Sequence: MW1|Offset1|MW2|Offset2|MW3|Offset3
Count
This value is always 1!
Temperature
Temperature in °C [max. 3 characters]
(char 2 digits, without decimal place)
In case isTemp=0, a „0“ is returned.
Example:
If you want to measure for example 20° and 85° angle without temperature with a three angle
gloss meter (20°/60°/85°), the following command string has to be used:
1| 401120999|xy|5|1|0:
Operation code:
1
(MeasureValue)
Serial-No:
401120999
(with preceding blank)
TID:
xy
(can freely be chosen)
AngleBinary:
5
(angle A1 and A3)
Count:
1
(this parameter is always 1!)
isTemp:
1
(additional temperature measurement)
The corresponding return-string could be as follows:
1| 401120999|xy|958|94|-1|-1|993|78|1|25