6.3.2 Command GetIsOnStandard
This command verifies if the measuring head is on the calibration standard.
GetIsOnStandard(SerialNo, TID)
OP-Code: 28
Parameter:
SerialNo
9-digit serial number with preceding blank
TID
2 characters transaction number of the host
Return string:
SerialNo
9-digit serial number with preceding blank
TID
2 characters transaction number
Flag
IsOnStandard
1 = is in the standard
0 = is not in the standard
6.3.3 Command MeasureTemp()
This command triggers a temperature measurement in the inner part of the measuring head.
Temperature variations can have an influence on the measured gloss value.
The function is not meant to determinate the absolute temperature but rather for the detection
of possible, bigger temperature differences, for example between the last calibration and the
precedent measurement.
MeasureTemp(SerialNo, TID)
OP-Code: 36
Parameter:
SerialNo
9-digit serial number with preceding blank
TID
2 characters transaction number of the host
Return string:
SerialNo
9-digit serial number with preceding blank
TID
2 characters transaction number
Temperatur
temperature value in °C
2 sign, no decimal place