
MG14 operating manual Rev 1.4
33
4.2.5.1 Degas menu
For each channel degas procedure can be activated ( for channel 2 only for ITR100, and ITR091 ).
Additionally degas time can be selected. Fig. 4-13 shows degas menu:
Fig. 4-13 Degas menu
4.2.5.2 Degas menu
–
Degas ON/OFF
For each channel degas procedure activation can be start independent. Using the narrow buttons
▲or▼ and MENU/ENTER button you can choose an option. Choosing Degas ON/OFF option the
following screen displays:
Fig. 4-14 Degas activation menu
Channel 2 degas activation is available only if ITR091 or ITR100 is set on this channel.
During degassing, pressure measurements cannot be performed. The measurement display shows
the
“Degasing …..” command.
The degas function is switched off automatically after the time declared in next submenu
– see
chapter 4.2.5.3. You may also deactivate this function manually at any time by pressing
MENU/ENTER button again.
During degassing procedure yellow LED on the front panel is illuminated too
– ref. to chapter 4.1.1.1.
NOTE:
There is no possible to start degassing procedure in both channels simultaneously.
Max. degas power which can be reach during the degassing procedure is set in the other menu
–
see chapter
Sensor
4.2.4.5.
Degassing power always increase with constant ramp
1W / s
, and can not be changed.