Pressure Systems, Inc. Model 9116 User’s Manual
Page 86
www.PressureSystems.com
The Valve Position indexes (
ii
=12 and
ii
=0C) each have two states (00/01) that when combined
provide four (4) possible states of the C1/C2/C3/C4 internal valves. This “Logical Rotary
Switch” with four (4) positions is summarized in the following chart:
C1 Energized
C2 Not
(0C=01)
C2 Energized
C1 Not
(0C=00)
C3 Energized
C4 Not
(12=01)
PURGE
position
LEAK/CHECK
position
C4 Energized
C3 Not
(12=00)
CAL/RE-ZERO
position
RUN
position
Example:
●
Send TCP/IP commands to
Model 9116
module (via its connected socket)
setting the calibration valve to the
CAL
(or
Re-Zero
) position:
“
w1200"
(Set
RUN/CAL
valve position)
“
w0C01
” (Set
CAL
position)
Responses
(both
commands):
“A”
“A”
Note
If the programmer “knows” that the module is already in the
RUN/CAL
valve mode, the first command above is optional. For more information,
see
Figures 4.1
through
4.4
in
Chapter 4
.