Pressure Systems, Inc. Model 9116 User’s Manual
Page 83
www.PressureSystems.com
SET/DO OPERATING OPTIONS/FUNCTIONS (Command ‘w’)
Purpose:
Change a module’s default operating option settings, or invoke special internal
operations or functions.
Command
“
wii[dd[ eeee]]
”
‘w’ is the
command
letter.
‘
ii
’ is the
option index
field.
‘
dd
’ is an [optional] hex
datum
field.
‘
eeee
’ is an [optional] extra datum with a leading
space
character.
Response
“
A
”
‘A’ is the
acknowledge
letter
Description:
The
index
field
(ii)
contains two hex digits that identify the specific option to be
set or function to be performed. The
datum
field
(dd)
, when present, contains 2
hex digits. A few indexes also require an
extra datum
field
(eeee)
. Valid
options/functions are listed in the table below (-- marks a missing
datum
field in
its column, and fourth column shows any
‘q’
command index that reads same
option):
ii
dd
Description
‘q’ read
index
00 ---
Execute
Internal Self Test
.
01 ---
Update
Internal Thermal Coefficients
.
02-06
---
Reserved for factory use
07 ---
Store
Operating Options
in non-volatile flash memory.
08 ---
Store
Current
Offset
Terms i
n transducers’ non-volatile
memories.
09 ---
Store
Current
Gain
Terms
in transducers’ non-volatile
memories.
0A 01-10
Set
Number of Channels in Module
(default =16 for 9116).
0B
00
01
Enable
Automatic Shifting of Calibration Valve
during
Calculate and Set Offsets
(
‘h’
) command (default).
Disable
Automatic Shifting of Calibration Valve
in
‘h’
.
User will manually control calibration value.