3.1.3 Responses
Four (4) types of responses can be returned from a
Model 9116
Intelligent Pressure
Scanner module:
•
an
Error
response,
•
an
Acknowledge
response,
•
an
Acknowledge with Data
response, or
•
a Network Query response.
The first three may be returned by TCP/IP commands, the latter from a UDP/IP command.
The error response consists of the letter ‘N’ (for NAK, or negative acknowledge), followed by a
2-digit hexadecimal error code. The following table lists the error codes that can be returned
from a
Model 9116
module:
Table 3.1
Error Codes
CODE
MEANING
00 (Unused)
01 Undefined
Command
Received
02
Unused by TCP/IP
03
Input Buffer Overrun
04
Invalid ASCII Character Received
05
Data Field Error
06
Unused by TCP/IP
07
Specified Limits Invalid
08
NetScanner error; invalid parameter
09
Insufficient source air to shift calibration valve
0A
Calibration valve not in requested position
The
Acknowledge
response is returned from a module when a command is received that
requires no data to be returned, and no error is detected. It indicates successful parsing and
execution of the last received command. It consists of the letter ‘A’ (for ACK, or acknowledge).
The
Acknowledge with Data
response is returned when a module receives a command
requesting data.
Model 9116
modules will typically return only the requested data values, each
preceded by a space character (except for format 7). No
‘A’
acknowledge letter begins this
data response.
Data are returned for the highest requested channel number first
.