Pressure Systems, Inc. Model 9116 User’s Manual
Page 25
www.PressureSystems.com
All modules effectively defer the host computer’s decision of “which channels of data do I want”
until that time when the host chooses to send read commands to actually retrieve the desired
data from the latest “buffered copy” of the continuously scanned, averaged, and engineering-
unit-converted data.
See
Section 3.1.4.4
(
Delivery of Acquired Data to Host
) for more information.
While scanning, all modules take multiple samples and average each channel. The
number of
samples per scanned channel
defaults to 8, but may be set to one (to disable averaging) or to
any power of 2 (1, 2, 4, 8, 16, 32) to change the degree of averaging (and its effect on
maximum scan rate). The
Set Operating Options
(
‘w’
) command may change this variable at
any time.
3.1.4.3
Calibration Adjustment of Offset/Gain Correction Coefficients
All
Model 9116
Intelligent Pressure Scanners have built in software commands (and pneumatic
hardware) to perform a periodic
zero
and
span
calibration adjustment of attached pressure
transducers. Use of these periodic adjustments result in the highest possible data accuracy.
The result of these calibrations are a new set of internal
offset
and
gain
coefficients. These
correction
coefficients are over and above those factory-determined and
unchanging
thermal
correction coefficients stored in each transducer's non-volatile memory. The factory coefficients
provide the basic engineering unit conversion capability, while also correcting for various non-
linear effects, including temperature effect compensation. The
offset
and
gain
correction
coefficients provide for fine linear fit adjustment of the factory calibration of each transducer.
If
used properly, the periodic zero and span calibration adjustment should be the only
calibration required to maintain specified performance through the life of the Intelligent
Pressure Scanner.
It is generally necessary for the transducer to have real
zero
and
span
pressure (specified as 2
or more values) applied when calibration adjustment is required. These pressure values may
be generated by secondary pressure standards, such as the model 903x calibrator module or by
other external means provided by the customer (such as a dead weight calibrator). For the
more common zero-only calibration adjustment, zero differential pressures can typically be
provided without the need for external pressure generators. All
Model 9116
modules have built-
in pneumatic inputs (CAL side inputs) and calibration manifolds required for directing the
generated pressures to the various channels of the module(s) being calibrated.
Refer to
Chapter 4 of this manual for detailed background and procedures for periodic calibration
of the Intelligent Pressure Scanners.
A summary of the commands used for calibration
purposes is included below.
The
Calculate and Set Offsets
(
‘h’
) command is executed only when a suitable “minimum”
(e.g., zero) pressure has been applied to all channels of the module. The new
offset
coefficients that result from execution of this command are stored in the module’s volatile (or
temporary) engineering-unit conversion database. They are also returned to the host in the
command’s response.
The
Calculate and Set Gains
(
‘Z’
) command should be executed only when “full-scale” (or
other suitable specified up-scale) pressure has been applied to the appropriate channels of a
module. The new
gain
coefficients that result from this command are stored in the module’s
volatile (or temporary) engineering-unit conversion database. They are also returned to the host
in the command’s response.
Содержание NetScanner 9116
Страница 136: ...Pressure Systems Inc Model 9116 User s Manual Page 127 www PressureSystems com Appendix C Cable Diagrams...
Страница 137: ...Pressure Systems Inc Model 9116 User s Manual Page 128 www PressureSystems com...
Страница 138: ...Pressure Systems Inc Model 9116 User s Manual Page 129 www PressureSystems com Appendix D 9116 Mounting Dimensions...