Pressure Systems, Inc. Model 9116 User’s Manual
Page 81
www.PressureSystems.com
cc
Transducer Coefficients Description
Datum
Type
00
Re-zero Cal Adjustment
(offset) term
(Note 1)
FLOAT
01
Span Cal Adjustment
(gain) term
(Note 2)
FLOAT
07
User Defined Field
(Note 4)
INTEGER
09
Transducer Manufacturing Reference Number
(Note 5)
INTEGER
0A
Transducer Full-Scale Range Code (See Appendix F)
(Note 5)
INTEGER
(Note 1)
Related command ‘
w08
’ can be used to download the offset term to the
sensor’s non-volatile memory (digitally or non-digitally compensated sensor).
(Note 2)
Related command ‘
w09
’ can be used to download the gain term to the
sensor’s non-volatile memory (digitally or non-digitally compensated sensor).
(Note 4)
Data is immediately stored to the sensor’s non-volatile memory.
The
User Defined Date
field (
cc
=07) is a 32-bit integer. Possible uses are to I
ndicate the date of last user zero and/or span calibration or possibly the date of
next required calibration. If this optional field is used, the user is responsible for
correctly encoding the date into the appropriate 32-bit integer value e.g., a
decimally encoded ‘yymmdd’ date. Any modifications of this field (using the
Download Internal Coefficients
(‘
v’
) command) will result in the new value
automatically being entered to transducer non-volatile memory.
A
special
single
Other Coefficients
array is selected with
array index
aa=
11
(hex).
All the valid
coefficient indexes
(for this array only) are listed in the
following
table:
cc
Other Coefficients Description
Datum
Type
01 EU
Pressure
Conversion scaler (default=1.0)
FLOAT