Pressure Systems, Inc. Model 9116 User’s Manual
Page 43
www.PressureSystems.com
READ TRANSDUCER A/D COUNTS (Command ‘a’)
Purpose:
Returns the most recently acquired
raw pressure
data for the specified channels
in averaged signed A/D counts (in the range -32768 to +32767). This simple
data bypasses any usage of the transducer’s factory-calculated coefficients or
the final calibration process’s adjustment coefficients (offset and gain). Each
datum returned in the response will be in the specified high-precision data
format, but representing A/D counts as a signed integer average. (The formula
for converting A/D counts to volts is:
Volts = A/D Counts * 5/32768
) This
command is intended for advanced users only and is not required for normal
operation.
Command
“
a
ppppf
”
‘a’ is the
command
letter
‘
pppp
’ is the
position
field
‘
f
’ is the
format
field
Response
“
dddd.. dddd
”
‘
dddd
’ are the
data
fields, each with leading space (except
f
= 7 or 8).
Description:
The 4-character hex
position
field (
pppp)
specifies a 16-bit binary bit-map, with
each bit (set to 1) specifying a particular channel number (16-1, left-to-right).
Only channels 12-1 are allowed for Models 9021 and 9022.
The 1-character
format
field (
f)
specifies the format of each
data
field (
dddd
) that
will be returned in the requested response. The first datum returned in the
response will be for the highest channel number requested. Each datum will be
preceded by a space character. Some formats may not be applicable to the
specific type of data being requested. Valid formats are shown in the following
table:
f
converts each internal response
datum
value from . . .
max. char.
0 single binary float
to 7-10-digit signed decimal “ [-xxx]x.xxxxxx”
13
1 single binary float
to 8-digit hex integer “ xxxxxxxx”
9
2 double binary float
to 16-digit hex integer “ xxxxxxxxxxxxxxxx”
17
5 single binary float
to long integer (EU*1000) then to 8-digit hex integer
9
7 single binary float
to single binary float (big endian: msb first)
4
8 single binary float
to single binary float (little endian: lsb first)
4