Pressure Systems, Inc. Model 9116 User’s Manual
Page 55
www.PressureSystems.com
Command ‘c’— Sub-command Index 03:
Clear Stream(s)
This sub-command is used to “undefine” any previously configured host stream in a module,
one at a time, or all together. The sub-command’s format is:
Command
“
c 03
st
”
‘c’ is the
command
letter
‘03’ is the sub-command index (‘
ii
’) for configuration
‘
st
’ is the stream identifier character (1, 2, or 3 or 0=
all streams
)
NOTE: all parameters are separated by a
space
character
.
Response
“
A
”
‘A’ is the
acknowledge
letter
Autonomous
Packet
none generated
Description:
This sub-command
clears
(un-defines) the particular specified host stream (
st
=1-
3), or un-defines “all configured” host streams with a single command (
st
=0).
Once cleared, a stream must be reconfigured before it can be started.
Example:
●
Stop all the streams configured previously. Then clear (un-define) only stream 3.
Finally, resume the remaining defined streams 1 and 2:
“
c 02 0
”
“
c 03 3
”
“
c 01 0
”
Read
response:
“A”
“A”
“A”