Pressure Systems, Inc. Model 9116 User’s Manual
Page 130
www.PressureSystems.com
Appendix E
Model 9116 Range Codes
The following range codes are stored in each DH200 pressure transducer. The range code of
each transducer can be read through the Read Internal Coefficient (‘u’) command.
Standard
Range Codes are shown in
Bold and Italics
.
Range Code
Full Scale Pressure
Minimum Calibration
Pressure
1
0.360 psi (10" Water Column)
-0.360 psig
2
0.720 psi (20" Water Column)
-0.720 psig
3
1 psid
-1.0 psig
4
2.5 psid
-2.5 psig
5
5 psid
-5 psig
6
10 psid
-5 psig
7
15 psid
-5 psig
8
30 psid
-5 psig
9
45 psia
0 psig
10
100 psia
0 psig
11
250 psia
0 psig
12
500 psia
0 psig
13
600 psia
0 psig
14
300 psia
0 psig
15
750 psia
0 psig
16
10 psid
-10 psig
17
15 psid
-12 psig
18
30 psid
-12 psig
19
45 psid
-12 psig
20
20 psid
-12 psig
21
20 psia
0 psig
22
15 psia
0 psig
Содержание NetScanner 9116