Pressure Systems, Inc. Model 9116 User’s Manual
Page 109
www.PressureSystems.com
Secure the PC-323 board in place using the long screw through the mounting bar and
the short screw into the hex standoff.
Figure 5.3c
PC-322 Board
(4)
Turn the assembly over
.
Replace the nylon washer and secure the nut to the
back of the long screw going through the mounting bar as depicted in
Figure 5.3c
.
(5)
Reassemble the scanner as previously described and test for proper operation.
5.1.4
Replacement of Transducers
Your
Model 9116
has
internal
DH200 pneumatic transducers, as well as an
internal
calibration
manifold with associated valves and O-rings. All these elements occasionally require service or
replacement as described in the following sections.
Following is a step-by-step procedure to replace a DH200 transducer in a
Model 9116
Intelligent Pressure Scanner. Use the tools and follow the general warnings already described
in
Section
5.1.1
.
Figure 5.4
Top View of DH200