Pressure Systems, Inc. Model 9116 User’s Manual
Page 73
www.PressureSystems.com
Example:
●
Send TCP/IP command to
Model 9116
module (via its connected socket), that
returns decimal pressure data for channels 1, 5, 9, and 13 in ASCII fixed point
format:
“
r11110
”
Response contains data for channels 13, 9, 5, and 1 (left to right):
“
1.234000 0.989500 1.005390 0.899602
”