73
-
Then, check and store the raw signal (this file could be demanded by PRECITEC technical
support team). The raw signal should show two peaks for channels which contribute to the
measurement.
Case 2: The target is positioned in front of the sensor, and thickness measurement is valid only for
some channels:
-
Set the measuring frequency to minimum ($SHZ100 or equivalent .NET DLL function).
-
Set the LED intensity to maximum ($LAI100 or equivalent .NET DLL function).
-
Check if the sample thickness variations along the line are fitting with optical head measuring
range; i.e. - Peak to Valley thickness variation should be inside minimal and maximal
measuring thickness range -, - maximum local slope on both target surface should be less
than maximum measuring slope -, - one or two target surface(s) are not homogeneous and
show high reflectivity variations along the line -
-
Set a mirror in place of your target and store the raw signal for each channels (this file could
be demanded by PRECITEC technical support team).
If after reading the previous section you didn’t succeed in resolving your problem, please contact your
vendor for technical support. In order to help to be more efficient, we recommend to fill the following
technical support datasheet and if necessary prepare the files which are demanded in previous
section.