71
Trouble Shooting
8.1
Power off:
The Status LEDs are off when the power supply cable is connected:
-
Check the CHRocodile CLS Switch button is ON.
-
Check the power cordon.
8.2
Communication error:
No possible Ethernet communication between CLS and your computer:
-
Check the Ethernet cable is plugged correctly in RJ45 sensor unit connector.
-
Check the IP configuration for your PC and the CHRocodile CLS unit (Cf. $IPCN command in
section 5.2.2).
-
If you are using a switch, then try to connect directly from your LAN network to the sensor unit.
-
If possible, a Peer-to-Peer connection is recommended.
8.3
Distance Measurement:
Case 1: The target is positioned in front of the sensor, and no distance measurement is collected for
all channels.
-
Set the measuring frequency to minimum ($SHZ100 or equivalent .NET DLL function).
-
Set the LED intensity to maximum ($LAI100 or equivalent .NET DLL function).
-
Check if a white light line emitted from the optical head is focused on the target.
-
Check if the target is inside the measuring range of the 192 channels (Cf. section 2.9.1).
-
Check if the target surface is normal to the optical head axis. The maximum angle between
target surface and optical axis is given by the maximum measurable slope (Cf. section 2.9.1).
-
Check if the measuring mode is Distance mode.
-
Check if the distance is transmitted ($SODX 16640 or equivalent function),