23
Signal
Pin
SyncIn
1
GND
2
SyncOut
3
Isolated-GND
4
[A] RS422 (CTS)
5
[A] RS422 (RTS)
6
[A] RS422 (RXD)
7
[A] RS422 (TXD)
8
Isolated-GND
9
Isolated-GND
10
Isolated-GND
11
[B] RS422 (CTS)
12
[B] RS422 (RTS)
13
[B] RS422 (RXD)
14
[B] RS422 (TXD)
15
Table 2.3: Interface (pin configuration)
Signal
Function
Description
RS432
RS422
Interface
Isolated (Isolated-GND) RS422 Interface
Sync In
Trigger-Input
Positive slope from 0V to 5-24V causes according to the
settings of the sensor:
starts the continuous measurement, if the command
wait for trigger was received first (TRG Command)
starts the single measurement in mode trigger each
(TRE Command)
the input is equipped with 10 kΩ pullup-resistance at
5 V.
Sync Out
Sync. Output
Isolated Sync Output
Positive slope 0 V to 5 V with the start of each measurement.
The pulse duration is 50µs in general and every second one
pulse is shorted to 10µs. This is for synchronizing more
sensors when their Sync-In is connected with the Sync-Out.
Table 2.4: Interface
Remark: As the Sync-input has a weak pull-up to 5V, your trigger source definitely needs to be able to
sink that current in order to pull the input down to Gnd. So as a trigger source, you can use an open
collector transistor output, that pulls to ground or a push pull output. The input can support 24V, but
the trigger threshold is always at approximately 2V. The trigger occurs (using the standard settings,