42
SEN
<0 .. 15, ?>
<optical probe Index>
„optical
probe“
index of used optical probe
SHZ
<32 .. 2000, ?> <x>Hz
x meaning the exact sample rate in
Hz in floating point format
“Set sample rate in Hz”.
SODX
<0..17> <0..17>
…
<0..17>
(max 16 times)
or <?>
<0..17> <0..17> … <0..17> (max 16
times)
“Set output data extended” definition of the output telegram
by enumeration of the indices of the data.
SSU
-
-
“Save Setup”, saves Setup Parameters to Eeprom Memory
STA
-
-
Start serial data output.
This mode will be stored in the Eeprom when executing the
SSU command. If stored the CHR will begin immediately to
output data telegrams on the next power up.
STO
-
-
Stop serial data output
This mode will be stored in the EEPROM when executing
the SSU command, so on the next power up the CHR will
not begin to send measurement data until the output is
restarted by the “STA” command
TABL
<Table
ID>
<args...>
<binary data,
?>
Upload table to device (or read back in case of '?'). See
below for details.
THR
<0..4094, ?>
<threshold value>
„threshold“ threshold for peak detection in the confocal
modes (0 and 1)
TRE
-
-
“Trigger Each” – Mode
TRG
-
-
”Wait For Trigger” Stops the sensor after completion of the
current data telegram and puts it in a waiting state.
VER
-
output version data
WHT
3141
Int.
ok![CR/LF]
or
Int.
too
weak![CR/LF]
or
Int. too high![CR/LF]
White Reference
Table 5.1: Commands list