31
Operational Start up
3.1
Connections and Interfaces
All of the connection ports are located at the rear of the CHRocodile CLS (Cf. section 2.6 Connections
and Interfaces):
the connection ports for the serial interfaces RS232/RS422
the connection for the encoder
the interface port for synchronization with external devices and
the power supply jack.
3.1.1
CHRocodile CLS Stand Alone device:
The device can be used as a stand-alone device in order to perform selective distance or thickness
measurements.
Power supply
The CHRocodile CLS has two pluggable screw terminal for power supply with 24V
DC
+/-10%.
Connect the set of power cable supply associated to the Power Supply Adapter (100-240VAC to
24V
DC
+/-10%) delivered with the CHRocodile CLS unit.
Ethernet connector
Connect the isolated RJ45 standard connector from the CHRocodile CLS unit to an Ethernet network
(PC). Use shielded cable for the data port connection (minimum category 5 cable).
The default CHRocodile CLS IP address is:
192.168.170.2
Configure the PC Ethernet port to the following address:
192.168.170.X (X≠2)
To configure the Ethernet port of your PC,
you must open the ‘
Network connection properties’
menu.
After selecting the right Ethernet card (connected to the sensor), click on ‘
network protocol (TCP/IPv4)’
and click ‘
Properties’.
Set the IP address of the PC and the mask. For a standard use the mask should
be set to 255.255.255.0.
If you need to configure the sensor to another IP address (different than
192.168.170.2)
, contact your
vendor. This could be useful in case of multiple CHRocodile CLS connection on 1 computer.
ON / OFF Switch button
Switch on the Power button.