37
In order to eliminate the influence of this undesirable light, a dark reference is performed on the sensor
(Cf command DRK in section 5). The Dark reference acquisition must be done when no object is in the
measurement range.
4.3
Auto Reference procedure
This procedure is no longer necessary. It has been disabled from CLS
Explorer version 1.0.1.1. The use of this procedure can modify the
calibration of your sensor, consequently if required, this procedure
must be done by PRECITEC, as a maintenance operation.
4.4
Mechanical interfacing
After completing the operational startup, i.e. connecting with the power supply, proceed with
initializing, then communication is ready and mechanical interfacing should be done.
Mechanical interfacing consists in:
-
Connecting the Optical Head which suit to your application to the CHRocodile CLS unit. The
Optical Head is simply screwed on the CHRocodile CLS unit (Cf. Figure 4.2),
-
And, fixing the CHRocodile CLS on your system using the interface threaded M4 holes (x11)
located on the soleplate of the Line sensor (Cf. section 6.3).
Fig 4-2: Interchangeable optical heads
Screw optical head
on sensor unit