15
3.
Remove the Fill adaptor Plug and closely inspect the port opening to ensure that no
debris, residue or moisture is present.
4.
Using a calibrated pressure gauge, check the supply cylinder to ensure that it contains at
least 2,700psi/ 186 BAR and less than 3,100psi / 214 BAR. It is very important to
ensure that the PEEBD1206is filled to its total capacity but not overfilled.
5.
Remove the protectivecap from the threaded nozzle of the fill adapter. Inspect the
nozzle to ensure the O-Ring is present and seated evenly at the base of the threads.
RED indicator visible below
the Control Knob, indicating
the “OFF” position
The RED indicator can be
easily viewed not matter the
position of the Control Knob