Cleaning, Disinfection & Sterilization Guide
Procedure 1
Cleaning and Sterilization Guide
for multi-part medical devices
Preliminary operations:
• Polydentia’s products are always supplied non-sterile.
For safe and optimal use always clean, disinfect and sterilize your product prior to use
• Check and inspect your equipment regularly.
Always refer to manufacturer’s instructions when using your equipment
• Please note that regulations may vary depending on the country where you are living.
Always refer to the regulation applicable in your country
• Materials may alter over time. Sterilization or exposure to chemicals can accelerate this deterioration.
Always
check your product for wear and damage before use
•
Avoid contact of dissimilar metals at any point during the cleaning and sterilization process
Cleaning:
Products should be cleaned as soon as possible after use to prevent soils from hardening
1. Disassemble all parts of the medical device
2. Remove visible debris or organic matter with a disposable wipe or surface brush.
This has to be done by scrubbing with warm tap water and soap or detergent
3. Inspect instruments for debris on surfaces. Repeat cleaning step 2 if there are still debris
4. Immediately soak in an appropriate disinfecting solution, following the manufacturer’s instructions
5.
Alternatively to point 4, an ultrasonic bath cleaner with an appropriate cleaning and disinfecting agent can be
used, in accordance with the manufacturer’s instructions. Please assure that the medical device has the relevant
symbol on the packaging
6. Rinse all parts with warm tap water
7.
Alternatively to points 2 to 6, a thermo disinfection with an appropriate cleaning agent can be used in
accordance with the manufacturer’s instructions. Please assure that the medical device has the relevant symbol
on the packaging
8.
Dry all the medical device components, then place them in an autoclave tray, or in a sterilization pouch (see
Sterilization instructions step 2 below)
Sterilization:
1.
Sterilization is not a substitute for proper cleaning. Clean the instrument fi rst (see section Cleaning) and
afterwards sterilize it
2.
Place all the medical device components in an autoclave tray, or in a sterilization pouch
3.
Sterilize in a steam autoclave using distilled water and at the temperature specifi ed on the relevant symbol
4. Do not use chemical autoclave or cold sterilization solutions
5. Avoid direct contact with the metallic part of the autoclave
IMPORTANT:
The medical device always has to be disassembled before reprocessing (cleaning, packaging, sterilization and storage).
DISCLAIMER:
This high quality product is manufactured by Polydentia SA, in Mezzovico, Switzerland. The content of this packaging has been
inspected, tested and approved by our quality offi
cers.
Materials used during our manufacturing process are designed to provide durability and performance. Polydentia constantly
looks for the highest quality plastics for its product; nevertheless please take into account that plastics may alter over time.
Sterilization or exposure to chemicals can accelerate this deterioration.
For the longest lifetime of your premium quality product always clean and sterilize it following these instructions.