Chapter
D - ADJUSTMENT & CALIBRATION
Planmeca Proline CC panoramic x-ray D-15
Technical Manual
Leave the ball phantom tool in position and place one end of the alignment ruler in the first pri-
mary slot (the largest one) and the other end in the secondary slot.
Figure
21
Loosen the four screws that hold the patient position mechanism in place and adjust its position
to the left or right until the mechanism is centered.
Figure
22
Make sure that the y-line on the ball phantom is parallel to the line on the underside of the align-
ment ruler. Note that the line on the ball phantom and the line on the alignment ruler do not have
to coincide, but they must be parallel, the deviation may be ±1mm.
Figure
23
Tighten the screws, and remove the alignment pin.
Take another ball phantom picture to check the alignment again. Readjust if necessary.
ST
AR
T
EN
D
PL
AN
ME
CA
No
.50
97
1
X-L
IN
E
Y-L
IN
E
Primary slot
Alignment ruler
Secondary slot
Screws
START
END
PLANMECA
No.50971
X-LINE
Y-LINE
S
Alignment ruler
Lines must not coincide
but they must be parallel
Ball phantom
Содержание Proline CC
Страница 2: ......
Страница 42: ...Chapter C HELP ERROR MESSAGES C 12 Planmeca Proline CC panoramic x ray Technical Manual...
Страница 158: ...Chapter E DIMAX3 DIGITAL SYSTEM ADJUSTMENT E 70 Planmeca Proline CC panoramic x ray Technical Manual...
Страница 186: ...Chapter H PARTS REPLACEMENT REPAIR H 8 Planmeca Proline CC panoramic x ray Technical Manual...
Страница 202: ...Chapter I FUNCTIONAL DESCRIPTION I 16 Planmeca Proline CC panoramic x ray Technical Manual...
Страница 203: ...Planmeca Proline CC panoramic x ray J 1 Chapter Technical Manual J SCHEMATICS DIAGRAMS...
Страница 204: ...Chapter J SCHEMATICS DIAGRAMS J 2 Planmeca Proline CC panoramic x ray Technical Manual...
Страница 206: ......