Chapter
G - TROUBLESHOOTING
Planmeca Proline CC panoramic x-ray G-13
Technical Manual
10 DISTURBANCES IN TEMPLE SUPPORT MECHANISM
10.1 Temple supports do not move
10.2 Temple supports function erroneously
SYMPTOM
POSSIBLE REASON
CORRECTIVE ACTION
Motor does not start
Keyboard out of order.
Motor out of order or connector
loose.
PCB out of order or connector
loose.
Motor starts but supports immobile
Pin loose.
Drawing nut glides on motor axle.
Thread of drawing nut damaged.
SYMPTOM
POSSIBLE REASON
CORRECTIVE ACTION
Temple supports do not close
properly
Attachment of quadrants to temple
support cranks has yielded.
Adjustment of magnet and Hall
sensor erroneous.
Temple supports do not move
symmetrically
Attachment of quadrants to temple
support cranks has yielded.
Temple supports have too much
clearance.
Adjustment of clearance incorrectly
done.