Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Planmeca Proline CC panoramic x-ray E-39
Technical Manual
6
CEPHALOSTAT ADJUSTMENT
NOTE
Switch the unit off after all the checks and adjustments have been performed, before tak-
ing any patient exposures.
6.1
Preparations before adjustments
Press the ceph mode selection key to enter the ceph mode if you are not already in it. The rotat-
ing unit will move automatically to the correct position for taking cephalometric exposures. The
latest used image area will be automatically selected.
Remove the inner cover from tube head assembly. Four screws hold the tube head cover in posi-
tion, these are located at the rear of the tube head assembly.
Figure
67
Rotate the cephalostat head support to the lateral ceph position. Remove the ear posts.
4
CEPH
Tube head
assembly
Tube head cover