Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Planmeca Proline CC panoramic x-ray E-37
Technical Manual
4.4
Adjusting the position of the optical sensor of the sensor head
NOTE
This section applies to pan/ceph units with two fixed sensor heads.
The panoramic position of the sensor head can be slightly adjusted by adjusting the optical sen-
sor position. Loosen the two sensor assembly attachment screws with the 2.5mm allen key and
move the assembly to desired position.
Figure
66
Optical sensor assembly