Chapter
G - TROUBLESHOOTING
G-10 Planmeca Proline CC panoramic x-ray
Technical Manual
7
DISTURBANCES IN CASSETTE MOVEMENT
7.1
Cassette carriage does not move
7.2
Cassette carriage stops suddenly
7.3
Trembles during movement
7.4
Abnormal noise from the mechanism
SYMPTOM
POSSIBLE REASON
CORRECTIVE ACTION
Cassette motor is not running
Keyboard out of order
PCB defective
Motor defective or connector loose
SYMPTOM
POSSIBLE REASON
CORRECTIVE ACTION
The motor stops running. If error
code appears, see error code
explanations.
Motor or PCB defective or motor
cable loose.
Motor stops/stalls
Gear assembly damaged.
One of the connectors loose.
Motor/drawing wheel glides
Linear bearing gets stuck.
Cassette carriage gets stuck to
secondary slot or another part of
the mechanism.
Grease on driving wheel or driving
surface.
SYMPTOM
POSSIBLE REASON
CORRECTIVE ACTION
Trembles during movement
Dirt in linear bearings or driving
surface.
Spiral gear damaged.
SYMPTOM
POSSIBLE REASON
CORRECTIVE ACTION
Abnormal noise from the mecha-
nism
Cassette carriage touches second-
ary slot or other constructions.
Inner secondary slot touches the
outer secondary slot in connection
with the step motor resonance.
Cassette glide/adhesive strips res-
onate.
Содержание Proline CC
Страница 2: ......
Страница 42: ...Chapter C HELP ERROR MESSAGES C 12 Planmeca Proline CC panoramic x ray Technical Manual...
Страница 158: ...Chapter E DIMAX3 DIGITAL SYSTEM ADJUSTMENT E 70 Planmeca Proline CC panoramic x ray Technical Manual...
Страница 186: ...Chapter H PARTS REPLACEMENT REPAIR H 8 Planmeca Proline CC panoramic x ray Technical Manual...
Страница 202: ...Chapter I FUNCTIONAL DESCRIPTION I 16 Planmeca Proline CC panoramic x ray Technical Manual...
Страница 203: ...Planmeca Proline CC panoramic x ray J 1 Chapter Technical Manual J SCHEMATICS DIAGRAMS...
Страница 204: ...Chapter J SCHEMATICS DIAGRAMS J 2 Planmeca Proline CC panoramic x ray Technical Manual...
Страница 206: ......