Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
E-22 Planmeca Proline CC panoramic x-ray
Technical Manual
The pmpcal2 window with text Waiting appears.
Figure
37
Press the ready key to drive the x-ray to the exposure position.
The text Exposure appears.
Figure
38
Take an exposure.
The sensor head is now calibrated. After the exposure has been taken the image is shown in the
Dimax2Tool window.
Figure
39
READY