Chapter
B - KEYBOARD FUNCTIONS & MODES
Planmeca Proline CC panoramic x-ray B-9
Technical Manual
Exposure time display
In the service mode this display shows the selected type of the X-ray tube. The codes corre-
spond to the following tube types:
Comet XL-90-5°/0.6Bfocal spot 0.6 X 0.6 mm
Toshiba D-052 SBfocal spot 0.5 X 0.5 mm
If the letters Fd is shown on the display, it indicates that the system has not been adjusted for the
filament parameters of the current tube and must be allowed to do this by pressing the exposure
switch down.
During exposure in the service mode this display shows the measured tube filament voltage.
Table 3: Meaning of indicator lights in the service mode
INDICATOR
LIGHT
NORMAL FUNCTION
SPECIAL FUNTION (CTL key is
activated)
The ready key indicator is blinking continuously
to indicate that the exposures can be started at
any time by operating the exposure switch.
The return key indicator is used to indicate the
primary slot limit sensor operation. When the
sensor is active (something in the sensor slot)
this light is on. When operated by the motor
movement the active time of the sensor is rather
short and is not necessarily always visible on
this indicator.
The sector A indicator is used to indicate the
temple rest motor limit sensor operation. When
the sensor is active (magnet on the sensor) this
light is on.
The sector A indicator is used to indicate the
cephalostat left cassette stop microswitch oper-
ation. When the sensor is active the light is on.
The sector B indicator is used to indicate the
layer adjust motor limit sensor operation. When
the sensor is active (magnet on the sensor) this
light is on.
The sector B indicator is used to indicate the
cephalostat right cassette stop microswitch
operation. When the sensor is active the light is
on.
The sector C indicator is used to indicate the up/
down motor upper limit sensor operation. When
the sensor is active (magnet on the sensor) this
light is on.
The sector C indicator is used to indicate the up/
down motor lower limit sensor operation. When
the sensor is active (magnet on the sensor) this
light is on.
The sector D indicator is used to indicate the
rotating arm start point limit sensor operation.
When the sensor is active (magnet on the sen-
sor) this light is on.
The sector D indicator is used to indicate the
secondary slot used. When the wide secondary
slot is selected, the indicator is on.
The sector E indicator is used to indicate the
rotating arm end point limit sensor operation.
When the sensor is active (magnet on the sen-
sor) this light is on.
The sector E indicator is used to indicate the
second up/down motor limit sensor.
The protrusive anterior arch indicator is used to
indicate the cassette holder start (ready) point
limit sensor operation. When the sensor is
active (magnet on the sensor) this light is on.
C.6
T.5
F d
READY
PRET
A
B
C
D
E
Содержание Proline CC
Страница 2: ......
Страница 42: ...Chapter C HELP ERROR MESSAGES C 12 Planmeca Proline CC panoramic x ray Technical Manual...
Страница 158: ...Chapter E DIMAX3 DIGITAL SYSTEM ADJUSTMENT E 70 Planmeca Proline CC panoramic x ray Technical Manual...
Страница 186: ...Chapter H PARTS REPLACEMENT REPAIR H 8 Planmeca Proline CC panoramic x ray Technical Manual...
Страница 202: ...Chapter I FUNCTIONAL DESCRIPTION I 16 Planmeca Proline CC panoramic x ray Technical Manual...
Страница 203: ...Planmeca Proline CC panoramic x ray J 1 Chapter Technical Manual J SCHEMATICS DIAGRAMS...
Страница 204: ...Chapter J SCHEMATICS DIAGRAMS J 2 Planmeca Proline CC panoramic x ray Technical Manual...
Страница 206: ......